Hotline: 0984.843.683 Email: info@ttech.vn  Zalo: 0984.843.683

5616 Secondary Reference PRT

Mã sản phẩm: 5616
Sử dụng cho Model:
Hãng SX: Fluke
Xuất xứ
Bảo hành
Tình trạng

Tình trạng: Còn hàng

Đặt hàng

Liên Hệ Hỗ Trợ

    • SĐT: 0984 843 683

    • Mail: info@ttech.vn

      • Zalo: 0984.843.683

Thông tin sản phẩm

Specifications

Parameter Value
Temperature range −200 °C to 420 °C
Nominal resistance at 0.01 °C 100 Ω ± 0.5 Ω
Temperature coefficient 0.003925 Ω/Ω/°C nominal
Calibrated Accuracy[1] (k=2)
 
 
± 0.012 °C at −200 °C
± 0.011 °C at 0 °C
± 0.028 °C at 420 °C
Short-term repeatability[2] ± 0.007 °C at 0.010 °C
Drift[3] ± 0.007 °C at 0.010 °C
Hysteresis ± 0.010 °C maximum
Sensor length 50.8 mm (2.0 in)
Sensor location
 
9.5 mm ± 3.2 mm from tip (0.375 in ± 0.125 in)
Sheath diameter tolerance ± 0.08 mm (± 0.003 in)
Sheath material INCONEL® 600
Minimum insulation resistance 500 MΩ at 23 °C
Transition junction temperature range[4] −50 °C to 150 °C (see footnote)
Minimum immersion length[5]
(< 5 mK error)
102 mm (4.0 in)
 
Maximum immersion length 254 mm (10 in)
Response time[5] 8 seconds typical
Self heating (in 0 °C bath) 60 mΩ/°C
Lead-wire cable type
 
 
PTFE-jacketed cable, PTFE insulated conductors, 24 AWG stranded, silver plated copper
Lead-wire length 182.9 cm ± 2.5 cm (72.0 in ± 1.0 in)
Lead-wire temperature range −50 °C to 150 °C
Calibration NIST-traceable calibration
[1]Includes calibration uncertainty and 100 hr drift.
[2]Three thermal cycles from min to max temp, includes hysteresis, 95 % confidence (k=2)
[3]After 100 hrs at max temp, 95 % confidence (k=2)
[4]Temperatures outside this range will cause irreparable damage. For best performance, transition junction should not be too hot to touch.
[5]Per ASTM E 644
 
 
 
 
 
 

Calibration Uncertainty

 
Temperature
Expanded
Uncertainty (k=2)
−197 °C 0.012 °C
−80 °C 0.012 °C
−38 °C 0.011 °C
0 °C 0.009 °C
156 °C 0.011 °C
230 °C 0.013 °C
420 °C 0.021 °C
Note: Laboratories may periodically reevaluate their uncertainties. Calibration uncertainties depend on the calibration process, the standards used, and the instrument performance.