Specifications
|
Resistance |
Nominal 100 Ω |
Temperature Coefficient |
0.00385 Ω/Ω/ °C nominal |
Temperature Range |
−200 °C to 420 °C (5627A-6 and 5627A-9 to 300 °C; transition and cable temperature: 0 °C to 150 °C) |
Drift Rate (k=2) |
± 0.04 °C at 0 °C after 100 hours at 420 °C |
Sheath Material |
316 Stainless Steel |
Leads |
PTFE-insulated, nickel-plated stranded copper, 22 AWG |
Termination |
Specify. See Ordering Information. |
Time Constant |
Four seconds maximum for 63.2 % response to step change in water moving at 3 fps. |
Bending Radius |
Sheath may be ordered with a bend on a minimum radius of 19 mm (3/4 in) except for 50 mm (2 in) area of sheath near tip. (Fluke lab requires 20 cm [8 in] of unbent sheath to re-calibrate.) |
Calibration |
Includes manufacturer's NVLAP-accredited (lab code 200706-0) calibration and table with R vs. T values in 1 °C increments from -196 °C to 500 °C (to 300 °C for 5627A-6 and 5627A-9).
ITS-90 coefficients included. |
Immersion |
At least 100 mm (4 in) recommended |
Calibrated Accuracy† (k=2) |
± 0.026 °C at -196 °C
± 0.046 °C at 0 °C
± 0.077 °C at 200 °C
± 0.124 °C at 420 °C |
Size |
5627A-12: 305 mm x 6.35 mm (12 in x 1/4 in)
5627A-9: 229 mm x 4.7 mm (9 in x 3/16 in)
5627A-6: 152 mm x 4.7 mm (6 in x 3/16 in) |
†Includes calibration uncertainty and 100 hr drift. |