General specifications |
Power requirements |
100 V ac to 240 V ac, 47 Hz to 63 Hz |
Fuse |
T2A 250 V ac |
Max power consumption |
100 W |
Operating ambient temperature range |
15 °C to 35 °C |
Storage temperature |
-20 °C to 70 °C |
Relative humidity |
Operating: <80 % to 30 °C, <70 % to 40 °C, <40 % to 50 °C |
Storage: <95 %, non-condensing. A power stabilization period of four days may be required after extended storage at high temperature and humidity. |
Vibration |
MIL-T-28800 |
Altitude (operation) |
<2000 m |
Ingress protection |
IEC 60529: IP20 |
Safety |
IEC 61010-1, Installation Category II, Pollution degree 2 |
Weight (Chassis only) |
13 kg (28.5 lbs) |
Dimensions |
Height: 147 mm (5.78 in) |
Width: 454 mm (17.79 in) |
Depth 488 mm (19.2 in) |
Rack mount dimensions |
3U-19 inch rack |
Warm up time |
15 minutes typical |
Control specifications |
Control precision (dynamic mode) |
PM200-BG2.5K range ± 0.005% range
All other ranges ± 0.001% range |
Control turndown1 |
10:1 (typical) |
Low control point |
1 kPa (0.15 psi) absolute |
Supply pressure requirements |
Clean dry N2 or air – Industrial grade nitrogen, 99.5 %+ |
Particulate contamination |
≤ 1.25 micrometer (50 microinches) |
Maximum moisture content |
-50 °C dew point |
Maximum hydrocarbon content |
30 ppm |