Specifications 9100S 9102S
Range 35 °C to 375 °C (95 °F to 707 °F –10 °C to 122 °C (14 °F to 252 °F) at23 °C ambient
Accuracy ± 0.25 °C at 50 °C; ± 0.25 °C at 100 °C; ± 0.5 °C at 375 °C ± 0.25 °C
Stability ± 0.07 °C at 50 °C; ± 0.1 °C at 100 °C; ± 0.3 °C at 375 °C ± 0.05 °C
Well-to-Well Uniformity ± 0.2 °C with sensors of similar size at equal depths within wells
Heating Times 35 °C to 375 °C: 9.5 minutes ambient to 100 °C: 10 minutes
Stabilization 5 minutes 7 minutes
Cooling Times 375 °C to 100 °C: 14 minutes ambient to 0 °C: 10 minutes
Well Depth 102 mm (4 in);1.6 mm (1/16 in) hole is 89 mm (3.5 in) deep 102 mm (4 in)
Removable Inserts; N/A Available in sizes from 1.6 mm (1/16 in) to 11.1 mm (7/16 in) [6.4 mm (1/4 in) and 4.8 mm (3/16 in) included]
Power 115 V ac (± 10 %), 55–65 Hz, 1.5 A or 230 V ac (± 10 %), 0.8 A, 45–55 Hz, 175 W 94-234 V ac (± 10 %), 50/60 Hz, 60 W;or 12 VDC
 
Size (HxWxD) 57 x 125 x 150 mm (2.25 x 4.9 x 6.1 in) 99 x 140 x 175 mm (3.9 x 5.5 x 6.9 in)
Weight 1 kg (2 lb 3 oz) 1.8 kg (4 lb)
Computer Interface RS-232 included with free Interface-it software
NIST-TraceableCalibration Data at 50 °C, 100 °C, 150 °C, 200 °C,250 °C, 300 °C, and 375 °C Data at –10 °C, 0 °C, 25 °C, 50 °C, 75 °C,100 °C, and 122 °C