Electrical Specifications |
Measurement function |
Alternating current |
AC ampere range |
3 mA, 30 mA, 300 mA, 3 A, 30 A, 60 A |
Range selection |
mA / A: manual selection
3 mA / 30 mA / 300 mA: automatic selection
3 A / 30 A / 60 A: automatic selection |
Range / Resolution |
3 mA¹ / 0.001 mA |
30 mA / 0.01 mA |
300 mA / 0.1 mA |
3 A / 0.001 A |
30 A / 0.01 A |
60 A / 0.1 A |
Accuracy – filter activated (40 to 70 Hz), filter off (40 to 1 kHz)² |
3 mA – 30 A
60 A |
1% + 5 digits
2% + 5 digits |
Frequency |
40 Hz to 1 kHz |
Crest Factor |
3 |
After calibration accuracy is valid for one year. Accuracy is expressed as ± (% of reading + digits). Reference conditions 23 ±5 °C and a maximum relative humidity of 80%.
¹The minimum value is 10 μA rms.
²Outside of the TC / °C 18 °C to 28 °C, 0.02 + 1 |
Physical |
Display (LCD) |
Digital readout: 3300 count |
Display refresh rate |
4 times / sec |
Maximum conductor diameter |
40 mm |
Dimensions |
234 x 101 x 46 mm |
Weight |
500 g |
Battery |
2 AA, IEC LR6, NEDA 15A, alkaline |
Battery life |
Expected battery life, without using backlight and spotlight, is more than 150 hours |
Automatic shut-down |
Meter automatically shuts down after 15 minutes of inactivity |