Sensors |
Up to four replaceable sensors: - LEL: Catalytic bead for combustibles (built-in Correction Factor library) - Oxygen:Liquid electrolyte O2 - Electro chemical sensor for Toxics |
Battery |
Rechargeable Li-ion |
Running time |
14 hourscontinuous non-wireless, diffusion 11 hourscontinuous non- wireless, pumped 10 hourscontinuous with wireless, diffusion 8 hourscontinuous with wireless, pumped Note: All battery specifications at 68° F (20° C); lower temperatures and alarm conditi |
Display Graphic |
Monochrome graphic display(128 x 80) |
|
Displaysize: 1.57" W x 1.06" H (40 x 27 mm) with backlighting |
|
Automatic or on-demand screen rotation |
Direct Readout |
Real-time reading of gasconcentrations |
|
Battery status |
|
Pump status(if equipped with pump) |
|
Wireless on/off and wireless reception quality |
|
STEL, TWA, peak, and minimum values |
|
Man Down and policy enforcement indicators |
Alarms |
Multi-tone 95dB buzzer (at 11.8"/ 30 cm, typical),vibration alarm, and flashing red LEDs and on-screen indication of alarm conditions |
|
Alarms: latching, non latching or manual override |
|
Additional diagnostic alarm and display message for low battery |
|
Pump stall alarm (pumped version only) |
|
Man Down Alarm with pre-alarm and real-time remote wireless notification |
Datalogging |
Continuous datalogging (3 monthsfor 4 sensors at 1-minute intervals) |
|
User-configurable datalogging intervals(from 1 to 3,600 seconds) |
Communication & Data Download |
Data download and instrument set-up and upgrades on PC via Travel Charger |
|
Wireless data and status transmission via built-in RF modem (optional) |
Wireless Network |
Mesh RAE Systems Dedicated Wireless Network |
Wireless Range (Typical) |
EchoView Host: LOS > 650 ft (200 m) |
|
ProRAE Guardian & RAEMesh Reader: LOS > 650 ft (200 m) |
|
ProRAE Guardian & RAELink3 Mesh: LOS > 330 ft (100 m) |
EMC/RFI |
EMC directive: 2004/108/EC |
|
EMI and ESD test: 100MHz to 1GHz 30V/m, no alarm |
|
Contact: ±4kV |
|
Air: ±8kV, no alarm |
IP Rating |
Pumped: IP-65 |
|
Diffusion: IP-67 |
Calibration |
Two-point calibration for zero and span (manual, or automatic with AutoRAE 2) |
Sampling Pump |
Built-in pump or diffusion Can sample through tubing up to 98ft (30m) |
Safety Certifications |
Europe: IECEx/ATEX (II 1G Exia IIC T4) |
Temperature |
-4°to 122°F(-20°to 50° C) for T4 temperature code |
Humidity |
0% to 95% relative humidity(non-condensing) |
Attachments |
Stainless-steel alligatorclip; Swivel beltclip (optional); Pouch (optional) |
Languages |
Arabic, Chinese, Czech, Dutch, English,French, German, Indonesian, Italian, Japanese, Korean, Norwegian, Polish, Portuguese, Russian, Spanish, Swedish and Turkish (language must be changed through ProRAE Studio II) |
Wireless Frequency |
ISM license free band. IEEE 802.15.4 Sub 1GHz |
Wireless Approvals |
FCC Part 15, CE R&TTE, Others |
Radio Module |
Supports RM900 |
Sensor Specifications |
|
Oxygen (O2) |
Range |
0 to 30.0% |
Resolution |
0.1% |
|
Combustible |
Range |
0 to 100% LEL |
Resolution |
1% LEL |
|
Carbon Monoxide (CO) |
Range |
0 to 500 ppm |
Resolution |
1 ppm |
|
Hydrogen Sulphide (H2S) |
Range |
0 to 100 ppm |
Resolution |
0.1 ppm |
|
Sulphur Dioxide (SO2) |
Range |
0 to 20 ppm |
Resolution |
0.1 ppm |
|
Hydrogen Cyanide (HCN) |
Range |
0 to 50 ppm |
Resolution |
0.2 ppm |
|
Ammonia (NH3) |
Range |
100 ppm |
Resolution |
1 ppm |
|
Phosphine (PH3) |
Range |
20 ppm |
Resolution |
0.01 ppm |
|
Chlorine (Cl2) |
Range |
0 to 50 ppm |
Resolution |
0.05 ppm |
|
Nitrogen Dioxide (NO2) |
Range |
0 to 50 ppm |
Resolution |
0.1 ppm |