Pressure Ranges
|
P3830 |
500 to 30,000 psi, or 40 to 2,000 bar, or 4 to
200 MPa, or 40 to 2000 kgf/cm2 |
P3840 |
500 to 40,000 psi, or 40 to 2,600 bar, or 4 to
260 MPa, or 40 to 2600 kgf/cm2 |
P3860 |
500 to 60,000 psi or 40 to 4,000 bar or 4 to
400 MPa or 40 to 4000 kgf/cm2 |
Accuracy
|
Accuracy* |
Standard accuracy is 0.02% of Reading*. Optional accuracy of 0.015% of Reading is available. ISO/IEC 17025 accredited calibration is provided with each instrument. *Accuracy is based on % of Reading from 10% to 100% of the piston range when used in accordance with the corrections found on the calibration certificate. Below 10%, ± (accuracy class) x 10% of the piston range. See the “Guide for the Uncertainty Analysis in Pressure when Using P3000 Series Deadweight Testers” technical note for more information. |
Materials of Construction
|
Standard weight material |
Series 3 non-magnetic, austenitic, stainless steel |
Weight density |
7.8 g/cm3 |
Piston Material |
Tungsten carbide with nickel binder |
Cylinder Material |
Tungsten Carbide with cobalt binder |
Thermal Coefficient of Expansion |
11 ppm/°C |
General
|
Test port adapters |
9/16 – 18 UNF (Autoclave); 3/8 BSP; 1/2 BSP & 3/4 BSP |
Instrument Weight |
66 lbs. (30 Kg) Instrument Base only |
Instrument Size |
18 in. X 13.5 in. X 19 in. (455 X 340 X 478 mm) |
Reservoir Volume |
235 cc (14.3 cu.in.) |
Intensifier ratio |
7 to 1 |
Seal Materials |
Nitrile |
Operating Fluid |
Dioctyl Sebacate (DOS), our reference 55-600 |
|