| Items for measurement | Total Organic Carbon | 
        
            | Measuring principle | Wet oxidization (UV Lamp) / NDIR measuring method | 
        
            | Measuring method | NPOC or TC-IC (alternativeness) | 
        
            | Measuring range | 0~5ppm, 0~10ppm, 0~100ppm, 0~1000ppm | 
        
            | Reaction time | Approx. 5 min. (90% reaction) | 
        
            | Channel | Max. six channels (optional) | 
        
            | Flow rate of sample water | Above 100mL/min., particle diameter <90um | 
        
            | Temperature of sample water | 5~50℃ | 
        
            | Resolution | 1ppb (for 0~5ppm) | 
        
            | Precision | ±2% of F.S | 
        
            | Acidic solution | Phosphoric acid solution (20%) | 
        
            | Oxidizing agent | Sodium persulfate | 
        
            | Calibration reagent | KHP 
 | 
        
            | Carrier gas | High-purity dry nitrogen : 0.2 ~ 0.8 MPa Consumption : approx. 80mL/min. for carrier, approx. 320mL/min. for aeration
 | 
        
            | Calibration | Manual(max. five point) or Automatic | 
        
            | Power | AC 220V, 50/60Hz | 
        
            | Display | 10.4 inch touch screen | 
        
            | Alarm indication on screen | No sample, no standard sample, no reagent, detector trouble, no carrier gas, transportation tube exchange, UV lamp exchange, calibration trouble, leakage, system alarm
 | 
        
            | Alarm output | TOC high/low measurement error, instrument trouble | 
        
            | Data output | 4~20mA (under 500Ω load resistance), RS-232 output | 
        
            | Data storage | Storage in PC and external USB memory | 
        
            | Applicable environment | 5~40℃ ambient temperature, 5~80% ambient humidity (no dew condensation) 
 
 |