Items for measurement |
Total Organic Carbon |
Measuring principle |
Wet oxidization (UV Lamp) / NDIR measuring method |
Measuring method |
NPOC or TC-IC (alternativeness) |
Measuring range |
0~5ppm, 0~10ppm, 0~100ppm, 0~1000ppm |
Reaction time |
Approx. 5 min. (90% reaction) |
Channel |
Max. six channels (optional) |
Flow rate of sample water |
Above 100mL/min., particle diameter <90um |
Temperature of sample water |
5~50℃ |
Resolution |
1ppb (for 0~5ppm) |
Precision |
±2% of F.S |
Acidic solution |
Phosphoric acid solution (20%) |
Oxidizing agent |
Sodium persulfate |
Calibration reagent |
KHP
|
Carrier gas |
High-purity dry nitrogen : 0.2 ~ 0.8 MPa
Consumption : approx. 80mL/min. for carrier, approx. 320mL/min. for aeration |
Calibration |
Manual(max. five point) or Automatic |
Power |
AC 220V, 50/60Hz |
Display |
10.4 inch touch screen |
Alarm indication on screen |
No sample, no standard sample, no reagent, detector trouble, no carrier gas, transportation
tube exchange, UV lamp exchange, calibration trouble, leakage, system alarm |
Alarm output |
TOC high/low measurement error, instrument trouble |
Data output |
4~20mA (under 500Ω load resistance), RS-232 output |
Data storage |
Storage in PC and external USB memory |
Applicable environment |
5~40℃ ambient temperature, 5~80% ambient humidity (no dew condensation)
|