Conductivity ranges
Accuracy
Resolution
|
0 to 200 µS/cm
0 to 2000 µS/cm
0 to 20 mS/cm
0 to 200 mS/cm
±0.5 % of range
0.1 µS/cm
1.0 µS/cm
10 µS/cm
0.1 mS/cm |
Temperature range
Accuracy
Resolution |
0 to 85 °C
±0.2 °C
0.1 °C |
Cell constant of sensors |
K = 1.0 1/cm |
Temperature coefficient of sensors |
According to EN 27888 (non-linear compensation) |
Temperature compensation |
Automatic |
Operating temperature
Device
Sensor |
0 to 50 °C
-5 to +80 °C |
Storage temperature |
-20 to +70 °C |
Voltage supply |
9 V block battery, type IEC 6F22 |
Current drawn |
Approx. 50 mA |
Special functions |
- Segment test at switch-on
- Measuring circuit monitoring
- "Hold" function (freeze latest measurement)
- MIN/MAX memory
- Automatic range changeover
- Automatic switch-off (can be de-activated)
- Battery change display
|
Dimensions
Hand-held meter
Sensor |
142 mm × 71 mm × 26 mm (L × W × H)
Length: 120 mm, diamter: 12 mm
Attached connecting cable, approx. 1 m long |
Housing material |
ABS |
Protection class |
IP65 at front |
Weight |
Approx. 225 g (with sensor) |