FeaturesDefib Surge P Resistor Network
Up to 380V test voltage.
Microcontroller technologies, front panel voltage meter indicates test peak open circuit voltage and polarity.
Output voltage and polarity are controlled manually from the front panel.
Ergonomically designed for safety, speed and efficiency.
Cables, manual and calibration certificate included.
External interlock disables HV output and defeats test when circuit is open.
Optional USB Testminder, computer control.
Optional Resistor network of Fig. 110.
One year calibration cycle.
One year warranty.
Specifications |
Output Voltage (Vtest) |
0 – 380 V +5%, -0% |
Rise Time (Tp) |
1.5mS ≤ Tp ≤ 2.5mS |
Duration (Tw50) |
3mS ≤ Tw50 ≤ 5.5mS |
Source Impedance |
10 Ω |
Rout |
R1 50 Ω ±5%, R7 5 Ω ±5% |
Voltage Control |
Manually, by turning a knob in front panel |
Polarity Control |
Positive and negative. Alternating control available optional computer control |
Voltage Display |
4 digit LED display |
Voltage Meter Accuracy |
±4V |
Duty Cycle |
1 pulse every 20 seconds |
Line Voltage |
120V AC, 50/60Hz *optional different line voltage |
Environmental |
Operating Temperature |
15-40 ºC |
Relative Humidity Range |
0-90% non-condensing |
General |
Dimensions |
17” wide x 3U high x 17” in deep |
Weight |
30 lbs approx. |
Product Package |
- Megapulse Defib Surge P 380 tester
- Megapulse Defib Surge P 380 product manual
- High Voltage Test Lead (Black), 4 feet
- Two High Voltage Test Leads (Red), 4 feet
- NIST traceable calibration certificate to ANSI Z540
- Calibration waveforms
|
Options |
XI |
External interlock; two position terminal block on the rear panel must be shorted for voltage operation. |
TMM |
Testminder USB computer interface software. The tester can be controlled with a computer. It can also be set to automatically follow test sequences and keep a record with test information. Compatible with Windows 32bit platform. |
220, 230 or 240 |
If a different line voltage is desired, with this option we install a step down transformer to be compatible with your line voltage. 220 is for 220V, 230 for 230V and 240 for 240V. |
Resistor Network |
Built in resistor network of Figure 110. Without the resistor network option the user must attach the remaining resistors outside the tester. |