Specifications |
ELECTRICAL |
Output |
Up to 7000 Volts at bulk cap; controlled by front panel knob. Bulk capacitor voltage is displayed on the front panel meter. Positive or ne gative polarity is chosen via front-panel button.
Virtual Impedance: 40 ohms. |
Voltage Waveform |
10x700us @ 1000-7kV at bulk cap: Per EN60950 Annex N (incl Australian Deviation).
Peak Value: 1000-7kV ±3% (independent of meter tole rance).
Front time: 10uSec ±30% [Trise= 1.67 (T90%-T30%)] per IEC 1180.
Duration: 50uSec ±20% (Time to half-value) per IEC 1180 |
Current Waveform |
40 ohm tester outputs up to 175A peak |
Meter Accuracy |
±3% 1000-7kV. |
Test Adjustments |
Amplitude via front panel knob and shown on front panel digital meter.Polarity adjustment via front panel button. |
ENVIRONMENTAL |
Operating Temperature |
15-40 ̊C. |
Relative Humidity Range |
0-90% non-condensing |
GENERAL |
Input Power Requirements |
114-128V, 50/60Hz, 2A. |
Weight |
30 lbs. approx. Shipping weight 40 lbs. |
Dimensions |
17 in. (W) x 3U in (H) x 17 in (D). 6U x 17" x 17" (Option IER: IER IEC Switch |
OPTIONS AND OPERATION |
Manual operation |
Operator connects the provided output and return c ables to the MegaPulse and the DUT. Operator select s waveform polarity, pushes the CHARGE button on the front panel and wa its until the front panel meter indicates test volt age. Operator conducts the test by pushing the TRIGGER button on the fron t panel. |
Options |
RI |
Relay isolated relays for connection to customer PLC via USB. |
TMM |
TestMinder allows test to be timed using computer clock and run from a PC. |
240 |
240V mains operation. |
230 |
230V mains operation. |
220 |
220V mains operation. |
BNCV |
Voltage BNC 1000:1 (Reference only). |
IER |
IEC 65 Figure 7b switch. (6U enclosure). |