Feature:
AC Output to 10,000 volts. Delivers up to 10 milliamps. Custom configurations up to 25 mA.
Logic controlled, one button test. LED’s indicate test progress. Upon failure, buzzer sounds and testing is terminated.
Leakage current limit, voltage ramp-up time and test duration are controlled from the rear panel. Output voltage adjustable from front and rear.
Automatic voltage rampdown circuit eliminates residual voltage.
Designed for product development and laboratory use.
High voltage test jack output.
Voltage ramp, shutdown on failure and test timer are defeatable.
1 year warranty
Output |
HT-10kVac |
to 10,000Volts ac. Leakage current 1 - 10mA AC. |
HT-10kVdc |
to 14,000Volts dc. Leakage current 1 - 5mA DC. |
HT-10kvac/dc |
combines outputs and adds a current meter to the front panel. 3U enclosure. |
|
Current capabilities of 25, 100 and 500mA available. Consult factory. |
Pass/Fail Criteria |
Leakage Current |
Adjustable via knob to any intermediate setting. |
Arc Detection |
Separate high frequency detection circuit for breakdown spike detection. |
User Adjustments |
Test Time |
1-60 sec., defeatable for operation while button is pressed. |
Voltage Ramp Time |
.5-5 seconds, defeatable. |
Leakage Fail Point |
AC - 1 - 10mA DC - 1-5mA via knob. |
Voltage Ramp Defeat |
applies voltage without ramp. |
Timer Duration Defeat |
applies voltage only while TEST button is pressed. |
Shutdown on Failure Defeat |
disables Arc Detection circuit. |
Operation |
Voltage rampdown time |
AC - .5 sec max; DC - 8 sec max. |
Duty Cycle |
100% |
Meter accuracy |
?3% 500-10kV |
Physical |
HT10kac and HT10kdc |
Weight 20 lbs. ac/dc: Weight 30 lbs. |
HT10kac and HT10kdc |
Dims 11.25Wx4.25Hx12D ac/dc 17Wx3UHx17D |
Input power requirements |
114-127VAC. |