Feature:
Performs BELCORE 1089 First & Second level AC power fault tests. (4.5.12, 4.5.13, 4.5.14, 4.5.15.1 and 4.5.16)
Performs UL 1950 Annex NAC powerline cross overvoltage tests.
Performs common mode (longitudinal) and differential mode (metallic) overvoltage tests.
Simulates contact between a power system primary and a telecommunications cable.
Simulates short term induction as a result of a power system primary fault to a multi-earth neutral.
Simulates long duration induction as a result of a power system fault to earth.
Single and dual channel models for 2 and 4 wire testing.
40 and 60 amp per channel models.
Preset and variable test times.
Adjustable source impedance.
Employs 0-volt switching for the most accurate repeatable test results.
Self contained in standard rack cabinet, casters included.
General |
Input Power Requirements: |
208 V, 2 conductor + ground, 120 A input current maximum for 1.5 seconds (100 A service OK) |
Output Voltage / Current Ranges: |
0-600V, 15Ω source impedance (for 600V, 40A test)
0-600V, 85.7Ω source impedance (for 600V, 7A test)
0-600V, 273Ω source impedance (for 600V, 2.2A test)
0-600V, 250-400Ω adjustable source impedance, 1A maximum
0-600V, 400-550Ω adjustable source impedance, 1A maximum
0-600V, 500-650Ω adjustable source impedance, 1A maximum
0-600V, 500-1150Ω adjustable source impedance, 0.25A maximum
0-600V, 1000-3750Ω adjustable source impedance, 0.5A maximum |
Weight: |
300 - 500 lbs. (model dependant) |
Dimensions: |
22”W x 19”D x 48”H |
Environmental: |
15-40°C operating temperature
0-90% Relative Humidity, non-condensing |