Features
Built specifically for the Test described in IEC 60950, Annex N, Table N.1, Line 1 and IEC 62368-1, Annex D, Table D.1, Line 1.
Charges the bulk capacitor to the required 2.5kV or 7kV (Australian Deviation), using the front panel knob or optionally via USB, see TMM-PF option.
Discharge of the 6838V output pulse is controlled by the front panel TEST button or optionally via USB, see TMM-PF option.
Meter reads the bulk capacitor voltage per the requirements of IEC 60950.
Polarity can be changed by the front panel POLARITY button or optionally via USB, see TMM-PF option.
Cables and NIST Calibration Certificate Included; A2LA Calibration optional.
One year warranty and unlimited telephone support.
Output Peak Voltage at 7kV: |
6838V ±3% |
Open Circuit Rise Time: |
9.13uS ±30% [Trise=1.67(T90%-T30%)] per IEC 1180 |
Virtual Impedance: |
40 ohms |
Open Circuit Duration Time: |
716uS ±20% (time to half-value) per IEC 1180 |
Current Waveform: |
Not defined |
Voltage Control: |
Manually, by turning a knob in front panel |
Polarity Control: |
Positive and Negative. Alternating control available wit optional computer control |
Voltage Display: |
4 digit LED display |
Voltage Meter Accuracy: |
±1% |
Charge Time: |
<15 seconds |
Line Voltage: |
120V AC, 50/60Hz |
Environmental |
Operating Temperature: |
15-40 ºC |
Relative Humidity Range: |
0-90% non-condensing |
Dimensions: |
17" wide x 3U high x 17" in deep. |
Weight: |
30lbs approx. (tester only 20lbs) |