Earth continuity
| Output Current |
200mA minimum into 2ohm |
| Test Voltage (o/c) |
9V DC nominal |
| Test Duration (Class I Test) |
2s nominal |
| Display Range |
0.01ohm to 19.99ohm |
| Pass Value (Class I Test) |
0.2ohm |
Insulation resistance
| Test Voltage |
250/500VDC @1mA nominal, < 2mA (s/c) |
| Test Duration (Class I or II) |
2s |
| Display Range |
0.01Mohm to 19.99Mohm |
| Pass Value |
1.0MOhm Class I |
| |
2.0MOhm Class II |
Equivalent leakage
| Test Voltage |
40V Nominal AC 50Hz Open Circuit |
| Test Duration |
2s |
| Display Range |
0.1mA to 19.99mA |
| Pass Value |
0.75mA/3.5mA (user selectable with 0.75mA as default)/0.25mA CL II |
IEC lead test
| Test voltage |
5V DC nominal |
| Test Duration |
Time for test (2s max) |
| Test |
Live/Neutral Checks for o/c, s/c and crossed |
Protective Conductor Current
| Test Voltage |
Supply Voltage, max current 13A |
| Test Duration |
120s maximum |
| Display Range |
0.15mA to 9.99mA |
| Pass Value |
0.75mA/3.5mA (user selectable with 0.75mA as default) |
Touch current
| Test Voltage |
Supply Voltage, max current 13A |
| Test Duration |
120s maximum |
| Display Range |
0.10mA to 3.50mA |
| Pass Value |
0.25mA |
3 phase differential leakage current
| *Requires 3 phase adaptor accessory |
|
| Test Voltage |
3 phase supply voltage |
| Test Duration |
Continuous measurement |
| Display Range |
0.10 mA to 9.99 mA |
| Pass Value |
0.75mA/3.5mA (user selectable with 0.75mA as default) |
RCD trip time
| Test Voltage |
Mains supply voltage |
| Test Current |
30mA/150mA rms sinusoida |
| Test Current Phase |
0°/180° auto switched on alternate tests |
| Measuring Range |
0ms – 500ms |
| Pass Value |
200ms |
Power socket check
| Input voltage range |
225 – 254VAC max current 13A UK |
| Indicates configuration of voltage potential: |
Line potential phase to earth
Line potential phase to neutral
Line potential neutral to earth |
|