Earth continuity
| Output Current |
200mA |
| Test Voltage (o/c) |
9V DC nominal |
| Test Duration (Class I Test) |
2s nominal |
| Display Range |
0.01ohm to 19.99ohm |
| Pass Value (Class I Test) |
0.2ohm |
Insulation resistance
| Test Voltage |
250/500VDC @1mA nominal, < 2mA (s/c) |
| Test Duration (Class I or II) |
2s nominal |
| Display Range |
0.01Mohm to 19.99Mohm |
| Pass Value |
1.0 Mohm Class I
2.0 Mohm Class II |
Equivalent leakage
| Test Voltage |
40V Nominal AC 50Hz Open Circuit |
| Test Duration |
2s nominal |
| Display Range |
0.1mA to 19.99mA |
| Pass Value |
0.75mA CLI / 0.25mA CL II |
IEC lead test
| Test voltage |
5V DC nominal |
| Test Duration |
Time for test (2s max) |
| Test |
Live / Neutral Checks for o/c, s/c and crossed |
Protective Conductor Current
| Test Voltage |
Supply Voltage, max current 13A |
| Test Duration |
30s maximum |
| Display Range |
0.15mA to 9.99mA |
| Pass Value |
0.75mA/3.5mA (user selectable with 0.75mA as default) |
Touch current
| Test Voltage |
Supply Voltage, max current 13A |
| Test Duration |
30s maximum |
| Display Range |
0.10mA to 5.00mA |
| Pass Value |
0.25mA |
3 phase differential leakage current
| *Requires 3 phase adaptor accessory |
|
| Test Voltage |
3 phase supply voltage |
| Test Duration |
continuous measurement |
| Display Range |
0.10mA to 9.99mA |
| Pass Value |
3.5mA |
RCD trip time
| Test Voltage |
Mains supply voltage |
| Test Current |
10mA / 30mA rms sinusoidal |
| Test Current Phase |
0°/180° auto switched on alternate tests |
| Measuring Range |
0ms - 500ms |
| Pass Value |
200ms
|
|