Technical Specification |
Measurement: |
4 Terminal Kelvin / Thomson principle eliminates errors due to lead resistance. |
Display: |
0.6in LCD 4000 count with automatic decimal point and polarity indication |
Ranges: |
6 push button selected with LED indication |
Terminals: |
0.2in binding posts accept spade tags and 4mm banana plugs |
Working temperature: |
0°C to +40°C re. humidity 80% max. non-condensing |
Storage Temperature: |
-20°C to +50°C |
Mains Supply: |
100/120/220/240V +10% to 13%
47Hz to 63Hz 20VA |
Safety: |
EN 61010-1 EMC - EN 61236 |
Dimensions: |
8.5in x 9.8mm x 0.3in (W H D) approx
1/2 19” rack 2U high
1.5ft x 14.73in x 8.5in packed in carton |
Mass: |
7.7lbs approx. 9.9lbs packed in carton |
Auto Zero: |
Permits the automatic zero of amplifiers and external circuits, eliminating errors due to thermal E.M.F. |
Protection: |
415 Vrms maximum at the measuring terminals will blow internal protection fuse |
Calibration: |
Digital security key protected |
Battery: |
6V 2.8Ah sealed lead acid battery with built-in charger. 14 hour typical operating time on lowest range, 28 hours typical on other rages. Internal charger and automatic switch off |
Resistance Measurement: |
4k Ω |
1 Ω |
100µA |
±(0.05% Rdg +0.02% FS) |
30ppm Rdg + 1ppm FS |
400 Ω |
100m Ω |
1mA |
± (0.05% Rdg +0.02% FS) |
30ppm Rdg + 1ppm FS |
40 Ω |
10m Ω |
10mA |
± (0.05% Rdg +0.02% FS) |
30ppm Rdg + 1ppm FS |
4 Ω |
1m Ω |
10mA |
± (0.05% Rdg +0.03% FS) |
30ppm Rdg + 4ppm FS |
400m Ω |
100µ Ω |
10mA |
± (0.05% Rdg +0.05% FS) |
30ppm Rdg + 25ppm FS |
40m Ω |
10µ Ω |
100mA |
± (0.05% Rdg +0.05% FS) |
30ppm Rdg + 25ppm FS |
|